Otii Ace Pro
Advance high-precision energy consumption analysis & optimization tool.
The Otii Ace Pro is the big brother of the Otii Arc Pro. Keeping the same portable and compact form with comprehensive technical features for energy optimization but with enhanced capabilities to boost battery-driven and energy-harvesting devices requiring high precision at higher voltage ranges, with a high sample rate and low step size.
An all-in-one tool with an isolated power supply unit, a constant voltage/current source and sink, and a high-precision multichannel multimeter. Making it a power analyzer/profiler that records and displays currents, voltages, and power measurements in real time. Enabling hardware, firmware, and software developers to optimize the energy consumption and battery life on any device under test.
With Otii Ace Pro and Otii 3 Desktop App, you can source the voltage of the device under test (DUT) up to 25V while simultaneously measuring and recording current, voltage, and power data in real-time for further analysis to optimize battery life throughout the development cycle. Supporting an adjustable sample rate up to 50ksps, plus a wide dynamic current measurement range (nA-5A) with 0.4nA resolution, and a current measurement accuracy of ±(0.05% + 25nA).
To take Otii Ace Pro's capabilities to the next level, and make it the perfect addition for product development, test and verification, quality assurance, and maintenance, explore Otii Toolbox for battery profiling and simulation or automation tools features.
Key benefits
All-in-one functionality: Otii hardware and Otii software integrates power supply, real-time monitoring, and detailed analysis in a single, compact, user-friendly product.
Quick installation & easy to use: Download Otii 3 Desktop app, plug the Otii Ace Pro to your device under test (DUT) and start measuring energy data in less than a minute.
Optimal power alternatives: Power up Otii Ace Pro based on your solution setup and power requirements via USB or an external DC adapter.
Versatile connection: Power source your device under test using the Otii Ace Pro, which can deliver up to 25V at high voltage/current resolution. The DUT can be connected either through its DC input or battery connectors.
Real-time measurements & analysis: Analyze and compare charts and measurements in real-time, covering currents, voltages, and power within the Otii 3 Desktop App – it allows scrolling, zooming, and selecting current consumption and debugging logs during measurement.
Continuous serial synchronization: Measurements are continuously synchronized with debug logs from the device under test via UART.
Expandable voltage & current: Connect multiple Otii Ace Pro in series to increase voltage and current capabilities to match your solutions needs.
Responsive UI: Ensure smooth performance under heavy data measurements loads, supporting multiple streams and long recordings – optimized for high-definition displays to enhance user experience during analysis.
Collaborative features: Share recordings and analyses with team members to get insights and feedback for further product releases on a recurring basis.
Upgradeable software: Unlock a wider range of features and specialized functionalities, including battery profiling and automation tools.
Key features
Hardware
Current and voltage measurement
Current measurement accuracy: ±(0.05% + 25nA) for -5A to 5A
0.4 nA current measurement resolution
24 bit ADC with automatic switching between ranges
No burden voltage
Voltage measurement accuracy ±(0.01% + 1 mV)
Connect multiple Otii Ace Pro in series to increase voltage and current capabilities to match your solution needs.
Sample rate
Adjustable sample rate up to 50 ksps for main current and voltage channel
Up to 50 ksps for all other channels (ADC current, ADC voltage, SENSE+, SENSE-)
Power supply
0V-25V
Isolated power supply, ±200V
-5A to 5A (depends on available current from USB or DC plug)
Digital interface
Digital IO voltage 1.2V - 5.0V
Software
Otii software support for Microsoft Windows, Apple macOS, and Linux Ubuntu.
Real-time graphs of currents, voltages, and power measurements and digital inputs.
Analyze statistics and measurements while recording continues in the background.
Connect multiple Otii Ace Pro to manage and sync multiple recordings at the same time.
Record and manage multiple recordings for unlimited time.
Record and sync data with UART logs.
Customize recordings and channels names to make the change logs easy to trace during your solution's development cycle.
Save, load, and export projects as a zip-archive, or recording data as CSV for further analysis in collaboration with your team.
In-line and 4-wire measurement support.
Unlimited changes with do/undo functionality.
Applications
Use Otii Ace Pro every day to:
Power your devices under development.
Energy profile microcontrollers, sensors, devices and electronics up to 25V in real-time and over time.
Optimize sleep current and extend battery life.
Measure inrush current.
Measure component leakage current over time.
Design power-efficient hardware, firmware and software through regression testing.
Troubleshoot your hardware and software.
Sync UART logs with power measurements to see what drains the energy
Datasheet
General
Temperature
10°C / 50°F
30°C / 86°F
Humidity
30 %
60 %
Main
Output voltage
0 V
25 V
Output voltage setting resolution
1 mV
Self-consumption
3.5 W
Output power, max continuous
30W ⁽¹⁻⁵⁾
Output power, max peak
50W ⁽¹⁾
Voltage between USB/DC jack and Main –
-200V ⁽²⁻³⁾
-200V ⁽²⁻³⁾
Voltage between DGND/AGND and Main –
-200V ⁽²⁻³⁾
-200V ⁽²⁻³⁾
Sink current
0 A
5 A
Sink current, setting resolution
1 µA
Sink voltage
0 V
25 V
Sink power, max continuous
15W
Sink power, max peak
125W
Accuracy -5A to 5 A
±(0.05% + 25nA)
Resolution
0.4nA
Internal sample rate
250ksps
Analog bandwidth (3 dB)
50kHz
Accuracy
±(0.01% + 1 mV)
Internal sample rate
250ksps
Analog bandwidth (3 dB)
50kHz
Output voltage readback resolution
3.5µV
Expansion port
Bitrate
50 bps
5.25 Mbps
Digital IO operating voltage
1.2 V
Vio ⁽⁷⁾
5 V
VIL Low-level input voltage
Vio*0.2V
VIH High-level input voltage
Vio*0.8V
Imax, Max sink/source current (total for GPIOs)
-10 mA
10 mA
Voltage input
-10 V
25 V
Shunt voltage range
-102.4 mV
102.4 mV
Resolution
12.2nV
Accuracy
±(0.1% + 1µV)
Input impedance
>100Mohm
Voltage input
-10 V
25 V
Resolution
3.1µV
Accuracy
±(0.1% + 250 µV)
Input impedance
>100Mohm
Voltage input
-10V
25 V
Resolution
3.1µV
Accuracy
±(0.1% + 250µV)
Input impedance
>100Mohm
Output voltage
0V
15V
Output voltage setting resolution
5mV
Output current
600mA
Voltage between USB/DC jack & DGND/AGND
-200V ⁽²⁻³⁾
-200V ⁽²⁻³⁾
USB and DC JACK
Input voltage
7 V
20 V
Input current
5 A
VBUS voltage ⁽⁴⁾
4.75 V
20 V
VBUS current ⁽⁴⁾
3 A
⁽¹⁾ Depends on available input power ⁽²⁾ USB and DC jack GND is connected internally to chassis GND ⁽³⁾ DGND and AGND are internally connected ⁽⁴⁾ USB PD 2.0 ⁽⁵⁾ Max 3A in on DC plug-in and max 4A output current